Standard reference platesKit HD0 - Additivity/Linearity
Set of two HD0 calibrated plates, containing UV filters in PMMA material, for carrying out the normative test to check theadditivity and linearity of your spectrophotometer in accordance with ISO 24443:2021, ISO/DIS 23675, ISO/CD 23698, FDA monograph 2011, etc.
Given their stable, standardized absorption and diffusion properties, they are suitable “reference emulsions” for checking and comparing instruments used for in vitro determination of UV protection for intralaboratory purposes.
absorbance (340 nm)
plates in kit
%
compliance with solar standards
Product information
Directions for use
Select two of the stable, standardized transparent PMMA UV reference plates. The absorbance peaks of these reference plates at 340 nm are at least 1.1 absorbance units, but less than 1.5 absorbance units (AU). Designate the first plate as plate A and the second plate as plate B.
- Measure against air,
- Place plate A in the light path of the spectrophotometer and measure in duplicate,
Place plate B in the light path of the spectrophotometer and measure in duplicate, - Place the two plates (A and B) one above the other with their rough sides facing each other in the light path and measure the combined absorbance (from 290 nm to 400 nm) in duplicate,
- Check that the minimum range limit is 2.2 AU and a linearity of at least 85%.
Technical information
Dimensions |
|
Weight | 4.5 g for one plate |
Packaging | Kit of 2 plates |
Compliance |
|
Optical property |
|
Chemical substances |
|
Conditions of use |
|
Storage |
|
Environmental features
Composition |
|
Made in France
Manufactured in an ISO 9001:2015 certified environment
Patented manufacturing